Hybrid Position and Force Control for Upper Limb Rehabilitation of Stroke Patient

Received: 21 May 2024     Accepted: 18 June 2024     Published: 2 July 2024
Views:       Downloads:
Abstract

This paper presents hybrid position and force control (HPFC) methodology for position and as well as force tracking for upper limb rehabilitation of stroke patients. Stroke is a leading cause of disability in humans. Traditional rehabilitative therapies help regain motor function and ameliorate impairment, but they depend on the therapist’s experience and require many therapists, which is cost-prohibitive. Most robotic tasks with high severity such as rehabilitation, demands an effective force as well as position control scheme to ensure the safe physical contact between the robot and its environment in this case the patient. Some rehabilitation robots have been developed to help stroke survivors recover motor function. In past robots used for upper-limb rehabilitation employed general control schemes such as proportional integral derivative. To overcome the problem of force tracking in rehabilitation, robots require modern control techniques. To mimic the human upper-limb, and universality of application, an end-effector based robot is used for this study. To ensure the convergence of position and force errors to zero extensive simulations are performed. Two of assistant modes passive and active assistive rehabilitation are considered. Most common rehabilitation trajectories horizontal reaching, and vertical reaching are selected as robot’s motion for both passive and active assistive activities. The mathematical model of robots’ kinematics, dynamics, alongside the proposed control scheme has been discussed in detail.

Published in American Journal of Mechanical and Industrial Engineering (Volume 9, Issue 2)
DOI 10.11648/j.ajmie.20240902.12
Page(s) 28-42
Creative Commons

This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited.

Copyright

Copyright © The Author(s), 2024. Published by Science Publishing Group

Keywords

Hybrid Position and Force Control, Position Tracking, Force Tracking, Rehabilitation, Stroke

1. Introduction
The interruption of the flow of blood to the brain is known as stroke which results in disfigurement of brain cells and can be untreatable. Stroke is among the leading and common causes of permanent disability with symptoms such as spasticity and the leading cause of mortality today.
It is undoubtedly the subsequent driving reason for death on the planet and one of the significant reasons for procured inability in grown-ups . Motor disfigurement on the upper limb and loss of hand agility are prime maladies that stroke tolerant have to suffer from, both of the above-mentioned dysfunctions are partially rectifiable by submitting to rehabilitation . Not only stroke many people suffer injuries of the upper limb frequently at extremities level such as fingers and hands followed by a forearm. Impaired motor function or any form of disability leaves much stress on families and much more pressure on patients themselves. Due to lack of technology, awareness, and facilities in hospitals the percentage of stable patients with reduced upper extremity who obtain dexterity is dangerously low. Current well-being administrations are grappling to provide indefectible rehabilitation therapy to stroke victims. This has roused scientists to investigate the utilization of robots to give restoration treatment to stroke patients . To fully recover motor impairment and joint mobility patients are required to execute exercises. The conventional method that has been used for decades for the rehabilitation of post-stroke is physiotherapy. Due to limited financial and human resources, physiotherapy does not promote recovery on an optimal level. The therapist to patient ratio in underdeveloped countries such as Pakistan is insolvent.
In the last three decades, the examination on the improvement of recovery robots has been increased exponentially, chiefly for the neurorehabilitation of post-stroke patients. Recovery has been demonstrated to be effectual when it is both acute and including for the patient , and robotics is one potential answer to impart intensity, by expanding the number of reiterations recommended by a specialist. The two significant highlights of the medicated exoskeletons are their capacities to apply forces alongside the assisted appendages and to give dependable joint estimations . Over the last two decades, force tracking remains the main concern of the researchers. Initially, the two major force control paradigms also known as classical force control schemes that have been proposed are 1) Hybrid force and position control and 2) Impedance control .
To address the joint mobility of upper extremity compound movement, rehabilitation robots are required to work on more than one DoF. Upper limb remediation robots can be halved into two major groups, exoskeleton, and end-effector-based. One of the earliest robots developed for upper limb rehabilitation is MIT-Manus which construct with a SCARA configuration and provides two translational DoF to advance the ante brachium and forearm . Extensive clinical testing of the MIT-Manus and the In-Motion ARM robot affirmed the enhancement of the motor range of the weakened arm after robotic treatment . In the second half of the most recent decade, various upper appendage restoration robots have been created. The first commercially available robot for upper limb rehabilitation came in 2011 .
The multiple degrees of freedom allow the robot to interact with the whole limb by applying distributed force. In the last decade, numerous rehabilitation robots have been made that use different control schemes. Two degrees of freedom (DoF) upper limb exoskeletal robot have been designed and produced which accompanies a proportional integral derivative (PID) controller for position control . A five DoF exoskeleton has been made which comprises impedance control to give the stroke victim the likelihood to effectively regulate the assignment and being passively steered by the robot when the victim is unable to complete that task . The local PID feedback controller along with impedance control is used to minimize the position error and force error between the human limb and robot . A visuo-vibrotactile feedback method is used on a planar robot for upper extremities rehabilitation . Artificial Neural Network (ANN) technique is used to directly predict the position angle for a three DoF underactuated robot arm to minimize the need for system modeling thus controller . For rehabilitation of elbow joint using resistance training a customized impedance and admittance control scheme is proposed that detects the motion of the human arm instead of the robot .
The force tracking error is unavoidable in all of the above methods. One of the basic ways to control the robot is to inactively control the movement along an ideal reference path using excessive gains. PID is a feedback control scheme similar to passive control, which governs position or force with respect to a known reference. Both of the above, mentioned methods are found to be useful only at the early stages of rehabilitation. A force control scheme that has the position signal as feedback is known as Impedance Control. While admittance control is opposite of impedance control where the position is being controlled while the feedback signal is in the form of force . Both are indirect force control methods. ANN technique opts for no force control because it is based on prediction and estimation of the control parameters as well as there is no dynamic relationship exist betwixt the human appendage and robot’s end-effector. The learning methods are not suitable for force tracking problems because these methods require a lot of data for training the system.
A lot of work has been done in designing the rehabilitation robot considering mechanical features such as inertia, friction, back drivability, etc. The key characteristics of rehabilitation robots lie in their control strategy as well. Only position control or force control strategy is not adequate for accurate implementation of upper limb rehabilitation, because the desired force should be preserved while traversing the prescribed path. To turn up the above drawbacks, the hybrid position and force control technique is developed and implemented , which states that for a given surface the end effector should be force controlled in the perpendicular orientation to the surface and position-controlled in the normal order to the surface.
Assistive mode provides rehabilitation in three forms known as passive, triggered passive, and partially/active assistive. The drawback of the passive technique is that it is only useful at the preliminary juncture of post-stroke healing. In this technique, we are only controlling the motion of the robot concerning a fixed reference path via position feedback control such as proportional-integral- derivative feedback (PID). In passive control, the reference trajectory is created by recording the physiotherapist’s inputs on a distorted limb which is a tiring process that can also contain human errors. In triggered passive control, the patient decides to trigger the assistance of a robot. As the patient is not participating in passive control and triggered passive control and all work is done by robots, their effectiveness for rehabilitation becomes questionable.
Assistance is found to be very encouraging in early patients as it introduces motivation as it allows the patient to actively control the motion. One of the decisive attributes for the meditated robot is to permit mutual control as soon as the stroke victim has recuperated least motor scope and articular surface mobility. This compliance between robot and patient cannot be achieved by using rigid controllers of high corrective gains. Instead of rigid controllers, modern rehabilitation robots rely on flexible control schemes such as Impedance control and HPFC.
Impedance control is an indirect force control strategy with position feedback. Apart from being an indirect force control scheme, impedance control is only coherent for lightweight rehabilitation robots. In rehabilitation robots, complications make an appearance where it is imperative to recoup for gravity and friction. On the contrary, the hybrid force, and position control scheme gives the vantage of including friction and gravity compensation. Upper limb rehabilitation demands that a driving force must be applied in a lengthwise direction concerning the covet track which guides the victim towards the precise implementation of the prescribed movement and a constraint force must be applied in a cross-over way to the motion to minimize errors. The HPFC scheme gives the freedom to independently control force and position simultaneously.
The rest of the paper is arranged as follows. Robot modeling that includes kinematics and dynamics are discussed in section 2. In section 3, robot control scheme and proposed approach is explained. Rehabilitation, and its types are discussed in section 4. Robot patient interaction is discussed in section 5. Simulations and results are presented in section 6 followed by conclusion in section 7.
2. Robot Modeling
In this chapter, we derived and explained the robot kinematics as well as robot dynamics formulation for the end effector-based robot used for rehabilitation. The kinematics of the robot describes the motion of the robot without considering the causes of the motion. When we are observing the kinematics of the robot, the torques and forces that cause the manipulanda proceed is not considered. On the other hand, the robot dynamics correlates the joint accelerations and joint torques.
2.1. Kinematics
The position kinematics can be subdivided into two categories, i) Forward Kinematics, ii) Inverse Kinematics.
2.1.1. Forward Kinematics
In the forward kinematics we have information about joint angles. And using joint angles we calculate end-effector location and orientation. While on the other hand in inverse kinematics the location of end-effector is known. We calculate joint angles based on the location of manipulanda.
In robotic applications, the DH-convention is utilized for frame allocation . The other method to calculate the forward kinematics is homogenous transformation. Using DH-convention, each link in the robot manipulator is assigned to a coordinate frame.
The homogeneous transformation of ith joint w.r.t (i − 1)th joint is mathematically represented as
Ti-1=i Rotz,θ.Transz,d.Transx,a.Rotx,a(1)
Tii-1=Rii-1Pii-101(2)
Tii-1=i-iiiiiii-ii00i0i0aiiaiidi1(3)
Where i, i, i, and i indicates cos(θ), sin(θ), sin(α), and cos(α) respectively. While Rii-1 and Pii-1 describes the rotation and translation of the i-1 joint w.r.t i joint.
Forward kinematics of 2R planar robot using D-H convention is mathematically represented as,
T02=A1.A2= c12-s120s12c120000010l1c1+l2c12l1s1+l2s1201 (4)
The frame of reference on each joint is represented as shown in figure 1. The DH parameters are presented in Table 1.
Table 1. DH parameters of 2 DoF Robot.

Link

Link Length

Alpha

Offeset

Theta

1

L1

0

0

θ1

2

L2

0

0

θ2

Below are the forward position kinematics equations for the 2R planar robot used for rehabilitation.
x= l1c1+l2c12(5)
y= l1s1+l2s12(6)
=θ1+θ2(7)
2.1.2. Inverse Kinematics
In inverse kinematics we used end-effector position to compute the joint angles. There are two approaches to derive the inverse kinematics, one is geometric, and other is algebraic. From the figure 1, applying Pythagoras theorem for joint angle θ2 gives.
Figure 1. Frame of reference on each joint of 2-DoF end-effector robot based on Denavit-Hartenberg convention.
x2+y2=l12+l22+2l1l2cosθ2 (8)
θ2=± atan sinθ2cosθ2(9)
Where,
cosθ2= 12l1l2x2+y2-l12-l22(10)
As
sinθ2=±1-cosθ22 (11)
Solving for θ1
θ1=atanl1 +l2 cosθ2y±(l2sinθ2)xl1 +l2 cosθ2x±(l2sinθ2)y (12)
Using the forward kinematics equations (5), (6), and (7), the correlation between the joint space and task space variables is expressed as
x=f(θ )(13)
The derivative of (13) w.r.t time gives,
ẋ=Jθ̇(14)
Where J describes the Jacobian matrix that establishes the relationship between joint velocities with the task-space velocities.
2.2. Robot Dynamics
The robot dynamics illustrates the relationship between the forces and motions. Given the joint torques, the joint accelerations computed by forward dynamics. On the other hand, inverse dynamics provides joint torques for given joint accelerations.
The two major approaches to develop the dynamics equations are Newton-Euler formulation, and Euler-Lagrange formulation. Newton-Euler method construct dynamic equations from forces and torques existing between each link in the manipulator. Whereas, Euler-Lagrange approach develop dynamic equations by examining the difference between Kinetic energy and Potential energy.
Using Euler-Lagrange formulation, the following dynamic equation of motion can be utilized to present the dynamic model of an n-link robot manipulator in the standard way.
Hqq̈+Cq,q̇q̇+gq+fq̇=T(15)
where,
q,q̇,q̈: Joint position, velocity, and acceleration
q ϵ Rn: Joint coordinate vector, n is number of joints
D, H(q): Positive-definite inertia matrix
g(q): Gravitational force
fq̇: Frictional forces
Cq,q̇q̇: Coriolis and centrifugal terms
τ: Joint torques
Equation of motion for the 2R planar robot in task space can be represented as
Hxẍ+Cx,ẋẋ+gx+fẋ=T(16)
where,
Hx=J-THJ-1(17)
Cxx , ẋ=J-TC-HJ-1 J̇ J-1(18)
gx=J-Tg(19)
fXẋ=J-Tf(20)
M1,1=(m1+m2)l12+m2l22+2m2l1l2c2(21)
M1,2=m2l22+m2l1l2c2(22)
M2,1=m2l22+m2l1l2c2(23)
M2,2=m2l22(24)
Mq̈=(m1+m2)l12+m2l22+2m2l1l2c2m2l22+m2l1l2c2m2l22+m2l1l2c2m2l22(25)
C1,2=-2m2l1l2s2θ̇1θ̇2-m2l1l2s2θ̇22(26)
C2,1=m2l1l2s2θ̇12(27)
Cq̇,q=-2m2l1l2s2θ̇1θ̇2-m2l1l2s2θ̇22m2l1l2s2θ̇12(28)
G1,1=(m1+m2)gl1s1+m2gl2s12(29)
G2,1=m2gl2s12(30)
Gq=(m1+m2)gl1s1+m2gl2s12m2gl2s12(31)
q=θ1θ2(32)
3. Robot Control
HPFC scheme is first proposed by J. Craig in 1981. As the impedance controller relies on the manipulator and external force being coupled together, in contrast in the HPFC scheme, the manipulator and end-effector force are not coupled together .
HPFC scheme is proposed so that both the position and force of the robot can be controlled simultaneously. Robotic tasks such as rehabilitation requires the robot to act in accordance with the environment, respond to the force applied on the end effector, and adapt to the position unpredictability of the environment. The architectural layout of the hybrid control is shown above in Figure 2.
Figure 2. Architecture Block Diagram of HPFC.
The elementary concept behind the formation of HPFC scheme lies in the task planning. Any desired task can be described with a set of constraints. With every key element of robotic task, a set of restriction are associated known as natural constraints. Natural constraints are formed due to the geometric limitations of the task. Artificial constraints come due to the desired behavior such as motion or force forms in task configuration.
Figure 3. Conceptual Block Diagram of HPFC.
Be mindful that the coordinate system that is used to specify constraints and trajectories does not necessarily correspond to the manipulator's joints or its end effector. Instead, the coordinates are selected in line with the task geometry. The conceptual diagram of the hybrid position and force controller is represented as shown in the Figure 3 above.
where,
Xd: Desired position (motion)
Xe: Position error
Fd: Desired force
Fe : Force error
S: Selection matrix
J: Jacobian
Xs: Sensed (actual) position
θe: Joint angle error
θs: Sensed (actual) joint angle
τ: Computed Torque
The above figure indicates that there are two control loops present int the hybrid control. The upper loop is of position control, and the lower loop is for force control. The force control will not be responsible for controlling the position of the robot. It will only make sure that the desired force is applied. Likewise, the position control will only be responsible for controlling the position parameter only. The position controller consider that the arm of the robot is in free space and no force is acting on it. The inverse jacobian is used to transform the position error from cartesian frame to joint error.
The basic premise of our minimal level hybrid controller is that each joint's actuator driving signal should represent that joint's instantaneous contribution to the fulfillment of the positional and force constraints.
The actuator driving signal can be expressed as,
τi=j=1MFijsjΔfj+Pij1-sjΔxj(33)
where,
τi: Torque applies by the ith actuator
Fij: Force compensator jth input and ith output
Pij: Position compensator jth input and ith output
Δfj: Force error in jth DoF
Δxj: Position error in jth DoF
sj : Component of compliance selection vector
sj=0 DoF under position control
sj=1 DoF under force control
The compliance selection vector S is a binary m-tuple. It distinguishes between the DoF under position control to the DoF under force control. The coordinate and kinematics blocks in the diagram indicate that the sensed position and force signal must be transformed into cartesian frame C attached with the task. That is implemented using the following equations,
XC=TJCθ(34)
FC=THCFH(35)
Where, {H} represents the hand frame, and {C} represents the Cartesian frame. For a given task the robot’s end effector should be position controlled normal to the surface, force controlled perpendicular to the surface. And the necessary criteria in order to implement is that there should be no friction present between the manipulators end effector and the task geometry or surface.
In the conceptual hybrid controller, besides error correction algorithms it is sensed that the ideal HPFC should include feed forward compensation for gravity, and robot dynamics.
To accomplish that hybrid control scheme is proposed in terms of joint space. The only assumption taken here is that the position of the task where a desired force is required to apply is known to a certain level, or there is an error margin available in the position of robot. This eliminates the transformation of position signal from Cartesian space to joint space, as well as force error signal. The proposed HPFC scheme is expressed as in Figure 4.
Figure 4. Hybrid Position and Force Control.
Where,
D(q): Inertia
G(q): Gravity Compensation
KF: Proportional gain for force controller
KPP: Proportional gain for position controller
KPV : Derivative gain for position controller
q̈ : Desired joint acceleration
The mathematical form of the joint space HPFC can be expressed as,
τ=τF+τP+G(36)
τF=JTFD+KFFD-F(37)
τP=Dq̈D+KPPqD-q+KPVq̇D-q̇ (38)
Where,
τF: Torque computed for force control
τP: Torque computed for position control
G: Gravity compensation
In terms of Cartesian space the hybrid controller with feed forward compensation that can also control the position and force independently is shown in the equations below. One of the major differences is that now the position control is comprised of two parts, position, and velocity. The control laws for the hybrid controller are expressed as
τP=KPPqet+KPIqetdt+KPDq̇et  (39)
τF=τfft+KfP τe t+KfIτe'tdt  (40)
Due to friction, and decay of damping with the speed, a non-linearity occurs. To overcome this non-linearity, the integral gain is used in the force control loop. To bound the integral gain values for non-linearity a saturation block is used.
4. Rehabilitation
The term rehabilitation indicates the process of restoring someone back to health or a regular life via training and therapy following incarceration, addiction, or disease. Neurorehabilitation is method for people who have any disorder in the nervous system, or faced injury. A general neurorehabilitation timeline for stroke patients is shown in Figure 5.
Figure 5. Neurorehabilitation Timeline.
Based on the time, the phases of rehabilitation can be divided into three categories known as acute phase, sub-acute phase, and chronic phase. From the time of occurrence to the first seven days, is called as acute phase. After the first week to the six months post stroke period is called as sub-acute phase. And after six months, it is called as chronic phase.
The type of neurorehabilitation strategy adapted for survivor is greatly dependent on the phase. In acute phase the limb is unresponsive, and it can hardly apply any force on the rehabilitative device. In sub-acute phase the limb is somewhat responsive and can apply minute force on the rehabilitation equipment. In the chronic phase the limb is responsive, and it can apply a considerable amount of force on the rehabilitative manipulator.
Globally the upper-limb robotic rehabilitation strategies for stroke patient are categorized into three domains i) Assistance, ii) Correction and iii) Resistance. The Figure 6 shows the neurorehabilitation strategies.
Figure 6. Global Strategies for Neurorehabilitation.
4.1. Assistance Mode
In assistance mode, the robot is assisting the patient by supporting the weight of the limb and by providing the forces to complete the task prescribed by the physiotherapist.
There are three strategies in assistive mode of rehabilitation named as passive control, triggered passive control, and partially assistive control.
4.1.1. Passive Control
One of the fundamental methods of controlling a manipulator is to control its motion against desired path through position feedback. In rehabilitation, the passive control is opted for early stages of post stroke therapy. Because in this phase the limb is totally unresponsive, so only the movement of the impaired limb through a prescribes path is considered. The major goal behind the passive control is that by exposing the impaired limb to the intensive iterative motion, the movement may be identified by the brain.
4.1.2. Triggered Passive Control
In this strategy, user selects when to activate the assistance of the robot. This technique is used where patient is interacting with the rehabilitation device using brain-machine interface.
4.1.3. Partially Assistive / Active Assistive Control
The major drawback of the passive control is that the patient is not actively contributing to the activity, because patient is not applying any force to perform the task. This reduces the required stimulation to affected muscles, and joints. As soon as patient recovered to a level where he can apply minimal force, the robot should allow the shared control. This scheme is known as active assistive control. Not only the shared coordination, the muscle strength can also be improved using this methodology.
4.2. Correction Mode
In correction mode, the robot is only providing the assistance when the patient is not performing the task accurately. On detection of deviation from the goal, the robot will force the injured limb towards the task.
4.3. Resistance Mode
In resistance mode, the robot works against the movements of the patient in order to increase the error. This will make the task difficult to achieve, and it will train the patient to adapt to unknown, and complex perturbations.
5. Robot Patient Interaction
One of the major goals of robotic-based rehabilitation is that robots are reprogrammable and multifunctional. These two attributes of robot allow it be used in different phases (acute, sub-acute, chronic) of rehabilitation. Also, for a particular patient a combination of different strategies (passive, active assistive) etc. found to be more effective in regaining the motor mobility.
5.1. Exoskeleton vs End-Effector Based Robot
There are two types of robotic devices used for upper limb rehabilitation, named as exoskeletons, and end-effector based robots. In terms of cost analysis, the exoskeletons are more expensive than the end-effector based robots.
One other major advantage the end-effector robots have over the exoskeleton is universality. As same end-effector based manipulator can be used for different patients, or for one patient in different phases without the need to modify hardware or software. On the other hand, a big advantage of exoskeleton robots is that, they can target the particular joint, or muscle of the patient as they apply distributed forces along the limb. Even though the exoskeleton robots are multi-contact systems, most of the rehabilitation robots are developed using end-effector manipulanda. In end-effector based manipulator, there is a single interaction point between the patient and the robot close to wrist but it provides motor training to the whole limb.
Figure 7. Robot Patient Interaction.
In this study we have taken the end-effector based approach as expressed in Figure 7. Two linkages that resemble the human upper arm and forearm shown in Figure 8 are part of the robot's design to mimic the human body (Figure 9).
Figure 8. End Effector Robot Model.
Figure 9. Posterior View of Human Upper Limb.
5.2. Patient Force
Depending on the patient's state, the nature of their deterioration, and the particular task or activity, the average force that a stroke victim applies might vary substantially. Following are some of the prominent aspects that contributes to the different forces. Severity of the stroke: (Mild/ moderate/ severe), Location of the stroke: (Cortical/ subcortical/ brainstem), Stage of recovery: (Acute/ subacute/ chronic), Type of movement (Grip/ arm extension/ shoulder flexion, etc.), and Individual's pre-stroke strength and function. Nonetheless, certain research offers valuable perspectives on stroke victims' ability to generate force:
A study on grip strength found that stroke patients had an average grip strength of around 10-20 N (Newtons) compared to 30-40 N for healthy individuals . According to a different study on arm motions, stroke patients' average forces during reaching and grasping activities ranged from 2 to 10 N .
5.3. Rehabilitation Trajectory
To encourage an adequate recovery, a planar robot intended for upper limb rehabilitation following a stroke ought to incorporate a range of training sessions incorporating various movement patterns. Horizontal reach , vertical reach, diagonal path, circular motion, and figure of eight are the most commonly implemented rehabilitative trajectories in rehabilitation manipulators. The first two horizontal and vertical reach are used as a robots’ prescribed path in this study and control parameters are fine-tuned accordingly.
Because the muscles utilized during each exercise vary, so do the effects of each approach on the patient. Horizontal reach found to improve shoulder flexion and abduction, as well as elbow and wrist movements.
While elbow extension also gets superior in the case of vertical reach in addition to shoulder abduction and flexion.
Deltoid, and rotator cuff muscles (Figure 10) are responsible for shoulder abduction and flexion, and stabilization. Biceps brachii and triceps brachii shown in (Figure 11) are responsible for flexion and extension movements (Figure 12) respectively . Brachioradialis of forearm, flexor and extensor of wrist and hand, and intrinsic hand muscles are common muscles that are triggered and actuated in both reaches.
Figure 10. Shoulder Muscles for Abduction, Adduction.
Figure 11. Elbow Joints Muscles for Flexion and Extension
Figure 12. Human Upper-Limb Joints Motions.
6. Simulations and Results
The proposed control scheme is implemented on the 2R planar robot to perform the passive, and active assistive upper limb rehabilitation. The real-world circumstances have been implemented into simulations to successfully validate the suggested strategy.
MATLAB and SIMULINK are used for performing a number of simulations based on rehabilitation applications. The full-arm dynamics have been incorporated within the simulations to examine real-time behavior. The robot’s model parameters are m1 = 2 kg, m2 = 0.850 kg, and links length are L1 = 35 cm, and L2 = 31 cm. The PID controller gain parameters are tuned using the SQP algorithm.
In passive rehabilitation, performance of the proposed control scheme is evaluated by observing the force (Figure 13) and position tracking errors (Figure 14). As we know, in passive rehabilitation the impaired limb is totally unresponsive. So, the stiffness observed by the robot is close to zero.
In active assistive rehabilitation, execution of the proposed control scheme is evaluated by observing the force and position tracking errors in both axes. Firstly, the position is controlled in x-axis, and force is controlled in y-axis. Then the position is controlled in y-axis while the force is being controlled in the x-axis. In active assisted rehabilitation the impaired limb is a bit responsive. So, the stiffness observed by the robot is taken as,
K=50(N/m), 0t10
For the force tracking in y-axis, and position in x-axis the steady state error is shown in Figure 15, Figure 16 respectively. Conversely, for the force tracking in x-axis, and position in y-axis the steady state error is shown in Figure 17, and Figure 18.
For the desired force, the torque applied by the patient on the end effector of the manipulator is shown in Figure 19.
Figure 13. Passive Force Tracking in Y-axis.
Figure 14. Passive Position Tracking in X-axis.
Figure 15. Active Assisted Force Tracking in Y-axis.
Figure 16. Active Assisted Position Tracking in X-axis.
Figure 17. Active Assisted Force Tracking in X-axis.
Figure 18. Active Assisted Position Tracking in Y-axis.
Figure 19. Torque Applied by Patient in Active Assistive Mode.
All the position, and force tracking graphs of active assisted rehabilitation indicates that both the position and force errors are converges to zero.
7. Discussion and Conclusions
In this study, HPFC scheme with optimized controller parameters is proposed to constructively deal with the force as well as position tracking problem. The proposed control structure traversed the desired trajectory while maintaining the desired force with the subject. In order to validate the proposed control scheme, two cases are considered: passive and active assistive rehabilitation. The proposed approach effectively resolved the force and position tracking problem in both cases of rehabilitation reasonably well. In addition, the proposed approach is able for time varying force tracking as well. The complete mathematical model of kinematics and full arm dynamics are presented herein along with the complete derivation of the proposed control scheme. Perhaps more crucial than its capacity to carefully monitor changes in reference input is the system's capacity to exert a consistent force on a moving environment. For multi degree of freedom position trajectories such as circle, the proposed approach will be tested on 3 DOF articulated robotic manipulator.
Abbreviations

HPFC

Hybrid Position and Force Control

DoF

Degrees of Freedom

DH

Denavit Hartenberg

PID

Proportional Integral Derivative

ANN

Artificial Neural Network

Acknowledgments
I would like to thank Human-Centered Robotics Lab team, and Principal Investigator Dr. Ali Raza for his support and guidance.
Author Contributions
Muhammad Ali: Conceptualization, Data Curation, Investigation, Methodology, Formal Analysis, writing – original draft, Resources
Muhammad Bilal: Data curation, Methodology, Investigation
Muhammad Nadeem Akram: Project Administration, Resources, Visualization
Mohsin Rizwan: Validation, Supervision, Visualization, Review
Funding
This work is not supported by any external funding.
Data Availability Statement
Not applicable.
Conflicts of Interest
The authors declare no conflicts of interest.
References
[1] V. S. Huang and J. W. Krakauer, “Robotic neurorehabilitation: a computational motor learning perspective,” vol. 13, pp. 1–13, 2009,
[2] C. L. Richards, F. Malouin, and S. Nadeau, Stroke rehabilitation: clinical picture, assessment, and therapeutic challenge, 1st ed., vol. 218. Elsevier B. V., 2015.
[3] H. S. Lo and S. Q. Xie, “Medical Engineering & Physics Exoskeleton robots for upper-limb rehabilitation: State of the art and future prospects,” Med. Eng. Phys., vol. 34, no. 3, pp. 261–268, 2012,
[4] M. Lotze, C. Braun, N. Birbaumer, S. Anders, and L. G. Cohen, “Motor learning elicited by voluntary drive,” pp. 866–872, 2003,
[5] J. J. Craig, “Hybrid Position / Force Control of,” J. Dyn. Syst. Meas. Control, vol. 102, no. June 1981, pp. 126–133, 1981.
[6] N. Hogan, “Impedance Control: An Approach to Manipulation: Part II — Implementation,” no. June 1983, 1985.
[7] H. I. Krebs et al., “Rehabilitation robotics: pilot trial of a spatial extension for,” vol. 15, pp. 1–15, 2004,
[8] B. T. Volpe, H. I. Krebs, N. Hogan, L. Edelsteinn, C. M. Diels, and M. L. Aisen, “Robot training enhanced motor outcome in patients with stroke maintained over 3 years,” Neurology, vol. 53, no. 8, pp. 1874 LP – 1874, Nov. 1999,
[9] R. Riener, U. Keller, A. Duschau-Wicke, V. Klamroth-Marganska, and T. Nef, “Transferring ARMin to the Clinics and Industry,” Top. Spinal Cord Inj. Rehabil., vol. 17, pp. 54–59, May 2011,
[10] A. Yavuz, M. E. Aktan, and A. T. Koru, “DESIGN, PRODUCE AND CONTROL OF A 2-DOF UPPER LIMB EXOSKELETAL ROBOT,” vol. 5, no. 2, pp. 119–130, 2019.
[11] A. Frisoli, F. Salsedo, M. Bergamasco, B. Rossi, and M. C. Carboncini, “A force-feedback exoskeleton for upper-limb rehabilitation in virtual reality,” Appl. Bionics Biomech., vol. 6, no. 2, pp. 115–126, 2009,
[12] B. C. Tsai, W. W. Wang, L. C. Hsu, L. C. Fu, and J. S. Lai, “An articulated rehabilitation robot for upper limb physiotherapy and training,” in IEEE/RSJ 2010 International Conference on Intelligent Robots and Systems, IROS 2010 - Conference Proceedings, 2010, pp. 1470–1475.
[13] G. Salas-lópez, O. Sandoval-gonzález, and I. Herrera-aguilar, “Design and development of a planar robot for upper extremities rehabilitation with visuo-vibrotactile feedback,” vol. 3, pp. 147–156, 2012,
[14] M. A. Fikri, S. C. Abdullah, and M. H. M. Ramli, “Arm exoskeleton for rehabilitation following stroke by learning algorithm prediction,” Procedia - Procedia Comput. Sci., vol. 42, pp. 357–364, 2014,
[15] Z. Song et al., “Implementation of resistance training using an upper-limb exoskeleton rehabilitation device for elbow joint,” J. Med. Biol. Eng., vol. 34, no. 2, pp. 188–196, 2014,
[16] T. Proietti, V. Crocher, A. Roby-Brami, and N. Jarrasse, “Upper-limb robotic exoskeletons for neurorehabilitation: A review on control strategies,” IEEE Rev. Biomed. Eng., vol. 9, pp. 4–14, 2016,
[17] G. N. S. D. L. Wedel, “AN EXPERIMENT IN HYBRID POSITION/FORCE CONTROL OF A SIX DOF REVOLUTE MANIPULATOR,”1988,
[18] J. Wiley, “Robot Modeling and Control”.
[19] N. Hogan, “Wp2 4: 00,” pp. 304–313.
[20] A. A. Maciejewski and C. A. Klein, “Obstacle Avoidance for Kinematically Redundant Manipulators in Dynamically Varying Environments,” Int. J. Rob. Res., vol. 4, no. 3, pp. 109–117, Sep. 1985,
[21] Bilal, Muhammad & Akram, Muhammad & Rizwan, Mohsin. (2022). Adaptive Variable Impedance Control for Multi-axis Force Tracking in Uncertain Environment Stiffness with Redundancy Exploitation. Control Engineering and Applied Informatics. 24. 35-45.
[22] Bilal, Muhammad. (2020). Implementation of Adaptive Variable Impedance Control for Dynamic Force Tracking on Redundant Manipulator.
[23] S. S Kim et al, (2017) Grip strength and hand function in patients with stroke.
[24] J. H. Cauraugh et al., (2005) Forces and moments generated by patients with stroke during reaching and grasping.
[25] Amador LDF, Castillo Castañeda E, Laribi MA, Carbone G. Design and Analysis of VARONE a Novel Passive Upper-Limb Exercising Device. Robotics. 2024; 13(2): 29.
[26] Direction-dependent differences in the quality and quantity of horizontal reaching in people after stroke, Shintaro Uehara, Akiko Yuasa, Kazuki Ushizawa, Shin Kitamura, Kotaro Yamazaki, Eri Otaka, and Yohei Otaka, Journal of Neurophysiology 2023 130: 4, 861-870.
Cite This Article
  • APA Style

    Ali, M., Bilal, M., Akram, M. N., Rizwan, M. (2024). Hybrid Position and Force Control for Upper Limb Rehabilitation of Stroke Patient. American Journal of Mechanical and Industrial Engineering, 9(2), 28-42. https://doi.org/10.11648/j.ajmie.20240902.12

    Copy | Download

    ACS Style

    Ali, M.; Bilal, M.; Akram, M. N.; Rizwan, M. Hybrid Position and Force Control for Upper Limb Rehabilitation of Stroke Patient. Am. J. Mech. Ind. Eng. 2024, 9(2), 28-42. doi: 10.11648/j.ajmie.20240902.12

    Copy | Download

    AMA Style

    Ali M, Bilal M, Akram MN, Rizwan M. Hybrid Position and Force Control for Upper Limb Rehabilitation of Stroke Patient. Am J Mech Ind Eng. 2024;9(2):28-42. doi: 10.11648/j.ajmie.20240902.12

    Copy | Download

  • @article{10.11648/j.ajmie.20240902.12,
      author = {Muhammad Ali and Muhammad Bilal and Muhammad Nadeem Akram and Mohsin Rizwan},
      title = {Hybrid Position and Force Control for Upper Limb Rehabilitation of Stroke Patient
    },
      journal = {American Journal of Mechanical and Industrial Engineering},
      volume = {9},
      number = {2},
      pages = {28-42},
      doi = {10.11648/j.ajmie.20240902.12},
      url = {https://doi.org/10.11648/j.ajmie.20240902.12},
      eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.ajmie.20240902.12},
      abstract = {This paper presents hybrid position and force control (HPFC) methodology for position and as well as force tracking for upper limb rehabilitation of stroke patients. Stroke is a leading cause of disability in humans. Traditional rehabilitative therapies help regain motor function and ameliorate impairment, but they depend on the therapist’s experience and require many therapists, which is cost-prohibitive. Most robotic tasks with high severity such as rehabilitation, demands an effective force as well as position control scheme to ensure the safe physical contact between the robot and its environment in this case the patient. Some rehabilitation robots have been developed to help stroke survivors recover motor function. In past robots used for upper-limb rehabilitation employed general control schemes such as proportional integral derivative. To overcome the problem of force tracking in rehabilitation, robots require modern control techniques. To mimic the human upper-limb, and universality of application, an end-effector based robot is used for this study. To ensure the convergence of position and force errors to zero extensive simulations are performed. Two of assistant modes passive and active assistive rehabilitation are considered. Most common rehabilitation trajectories horizontal reaching, and vertical reaching are selected as robot’s motion for both passive and active assistive activities. The mathematical model of robots’ kinematics, dynamics, alongside the proposed control scheme has been discussed in detail.
    },
     year = {2024}
    }
    

    Copy | Download

  • TY  - JOUR
    T1  - Hybrid Position and Force Control for Upper Limb Rehabilitation of Stroke Patient
    
    AU  - Muhammad Ali
    AU  - Muhammad Bilal
    AU  - Muhammad Nadeem Akram
    AU  - Mohsin Rizwan
    Y1  - 2024/07/02
    PY  - 2024
    N1  - https://doi.org/10.11648/j.ajmie.20240902.12
    DO  - 10.11648/j.ajmie.20240902.12
    T2  - American Journal of Mechanical and Industrial Engineering
    JF  - American Journal of Mechanical and Industrial Engineering
    JO  - American Journal of Mechanical and Industrial Engineering
    SP  - 28
    EP  - 42
    PB  - Science Publishing Group
    SN  - 2575-6060
    UR  - https://doi.org/10.11648/j.ajmie.20240902.12
    AB  - This paper presents hybrid position and force control (HPFC) methodology for position and as well as force tracking for upper limb rehabilitation of stroke patients. Stroke is a leading cause of disability in humans. Traditional rehabilitative therapies help regain motor function and ameliorate impairment, but they depend on the therapist’s experience and require many therapists, which is cost-prohibitive. Most robotic tasks with high severity such as rehabilitation, demands an effective force as well as position control scheme to ensure the safe physical contact between the robot and its environment in this case the patient. Some rehabilitation robots have been developed to help stroke survivors recover motor function. In past robots used for upper-limb rehabilitation employed general control schemes such as proportional integral derivative. To overcome the problem of force tracking in rehabilitation, robots require modern control techniques. To mimic the human upper-limb, and universality of application, an end-effector based robot is used for this study. To ensure the convergence of position and force errors to zero extensive simulations are performed. Two of assistant modes passive and active assistive rehabilitation are considered. Most common rehabilitation trajectories horizontal reaching, and vertical reaching are selected as robot’s motion for both passive and active assistive activities. The mathematical model of robots’ kinematics, dynamics, alongside the proposed control scheme has been discussed in detail.
    
    VL  - 9
    IS  - 2
    ER  - 

    Copy | Download

Author Information
  • Department of Mechatronics & Control Engineering, University of Engineering and Technology, Lahore, Pakistan

    Biography: Muhammad Ali did his BSc in mechatronics engineering from Air University Islamabad, and MSc in mechatronics engineering from University of Engineering and Technology (UET) Lahore. He has worked as a mechatronics engineer at Epazz Inc. Currently he is serving as an embedded quality engineer at Powersoft19. Prior to that he has worked as an Instructor at the Human-Centered Robotics Lab, Al-Khawarizmi Institute of Computer Science, University of Engineering and Technology (UET), Lahore, Pakistan.

    Research Fields: Rehabilitation Robots, Force Control, Human Robot Interaction, Safety Critical Embedded Systems

  • School of Computing and Information Systems, University of Melbourne, Melbourne, Australia

    Biography: Muhammad Bilal did his BSc in mechatronics engineering from UET, Faisalabad, and MSc in mechatronics engineering from UET, Lahore. He has worked as a Research Officer at Human-Centered Robotics Lab, Al-Khawarizmi Institute of Computer Science, University of Engineering and Technology (UET), Lahore, Pakistan, 2019-2022. He worked as Graduate Assistant at the Mechatronics department, UET, Lahore for six months. He has participated in multiple international research collaboration projects in recent years. He is currently pursuing his PhD in robotics from School of Computing and Information Systems, University of Melbourne, Australia.

    Research Fields: Human-Robot Interaction, ROS-Moveit, LLMs, Robot Calligraphy, Human Computer Interaction

  • Department of Mechanical, Automotive, and Materials Engineering, University of Windsor, Windsor, Canada

    Biography: Muhammad Nadeem Akram is a PhD student of Industrial and Manufacturing System Engineering in the Department of Mechanical, Automotive, and Materials Engineering, Faculty of Engineering, University of Windsor, Windsor, Ontario, Canada. He earned his B.Sc. In Mathematics and Physics from the University of Punjab, Pakistan, his B.Sc. in Electrical Engineering from the University of Engineering and Technology, Pakistan, his Master of Industrial Engineering, and his Master of Applied Sciences in Industrial Engineering from the University of Windsor, Windsor, Ontario, Canada.

    Research Fields: Robot Design, Control Design, BMS, EMS, Sustainable EOL EVs

  • Department of Mechatronics & Control Engineering, University of Engineering and Technology, Human-Centered Robotics Lab, Lahore, Pakistan

    Biography: Mohsin Rizwan did his BSs in mechanical engineering from UET Lahore and did his MS and PhD from University of Texas at Arlington in 2011. He served UT Arlington as Teaching Fellow and research associate and joined General Electric in 2012. He left General Electric and joined UET Lahore as Assistant Professor in 2013. Currently he is serving Mechatronics Department of UET as Associate Professor. He has won three research grants. He is Co-PI of the Human-Centered Robotics Lab of the NCRA with a focus on the development of active/passive exoskeletons for industrial & biomedical applications.

    Research Fields: Assistive Robots and Prosthesis, Control Design and Optimization, Development of Model-based control, System-level mathematical modeling, Response analysis of mechanical/electromechanical systems